Repeater RIP 60 HA


Download PDF

Receiving data from RS232, RS485, RF 868 MHz.

Numerical Display to 6 characters/figures, height 60 mm.

Container in stainless steel and watertight logons.

Connects to all instruments of our production.

Configuration and customization protocol with PC software supplied.

Download installation manual

Download usage manual


Zoom Click to enlarge image

Repeater RIP 60 HA

Technical Specifications

Power supply230 - Opz. 115 Vac 50/60 Hz
Max power consumption15 VA
Operating temperature-10 / +40°C (14 °F / 104°F) Humidity maximum 85% non-condensing
Storage temperature-20 / +50°C (-4 / +122°F)
DisplayHigh efficiency 6-digit numerical LED and wide viewing angle, reading distance up to 50 feet
 Reading distanceTill at 15 m
 Visualized dataWeight acquired from cells or analogue input 0-10V/4-20 mA, alphanumeric indications/programming menu
 Functions remote controlTare entry / cancellation, zero scale, serial transmission current weight, configuration and calibration
Overall dimensions320 x 110 x 90mm / 12.6 x 4.3 x 3.5 in (l x h x p)
MountingWall-mounted with 4 external side holes Ø 6 mm with center distance 340 x 80 mm (l x h)
Case materialStainless steel protected against dust
Protection levelIP 65
ConnectionsWaterproof circular connectors with conductor wiring with screw terminal
Communication portsRs232c / Rs485 standard
 Max. cable length15m (Rs232c) and 1000m (Rs485)
 Serial protocolsASCII + programmable custom protocols
 Baud rateFrom 1200 to 115200 selectable
 OptionsEthernet protocol and Profibus protocol
Optional interface RF 433 MHzCoupled operation with transmitter module to be combined with the transmitter weighing instrument
Num. RF ChannelsUp to 7 channels selectable on the 868 MHz band for installation of multiple systems in the same area
Average radius of coverage100 meters in free air
Conforms to standardsEN61000-6-2, EN61000-6-3, EN61010-1
For the complete and latest technical specifications, please refer to the technical data sheet which can be downloaded here:
Download PDF

Connections

Type9P D-Sub J2 RIP Connector
LengthND
1RS485 -
2RS232 Tx
3RS232 Rx
4NC
5GND
6NC
7NC
8NC
9RS485 +

MOUNTING OF THE LOAD CELL

General precautions to be followed in the installation and assembly of load cells:

  • Follow the direction of application of force on the load cell
  • Do not exceed the limits of the temperature as indicated in the data sheet
  • Do not remove labels from the load cell
  • Adequately protect the cables with sleeves or channels
  • Carry out on each load cell a ground connection with suitable cable, shunt on mounting accessories
  • Do not carry out any welding on the mechanical structure after mounting the load cell
  • Avoid or reduce to a minimum level any mechanical constraints between the rigid structure and the weighing structure to avoid error in weighing and return to zero
  • In case of vibration or dynamic loads product the load cells with puffers / shock absorbers
  • Protect load cell from any dust accumulation with any appropriate means